Test item | Test summary | Probability component P |
Dose combinations per pen R |
Total measurements per V measurement Number of measurements n |
Bilateral target K-value |
Test basic principle (Refer to YY/T1768.1-2021 standard) |
Cool, standard and warm atmospheres | 5℃±3°℃、23℃C±5℃ and 40°±2℃ test dose accuracy (DA) | 0.975 | 1 | 60 | 2.670 | These tests are expected to measure the performance of the pen injector over a range of temperature and humidity variations representative of indoor or outdoor controlled or uncontrolled "in-use" environments. Outdoor conditions include seasonal (winter to summer) variations and indoor conditions include year-round variations in "room temperature". |
Free Fall | Test dose accuracy (DA) after a one-meter fall in three directions | 0.95 | 1 | 21 | 2.731 | The purpose of the free fall test is to verify the performance of the product after landing on impact in the unused condition without the outer packaging (or carrying case). A nominal height of 1m is used (this is the most likely height at which the pen injector will be placed on a table). |
Dry Heat/Freezing | 70℃±2℃ and -40℃±3℃,test dose accuracy(DA) | 0.975 | 1 | 60 | 2.670 | The test is expected to measure the performance of pen injectors after exposure to extremely hot and cold storage and transportation conditions, as well as extreme behaviors with potential users (e.g., placing a pen injector on the dashboard of a car in hot weather or accidentally storing it in a freezer). |
Vibration | Test DA after vibration | 0.975 | 1 | 20 | 2.760 | The vibration test is expected to simulate ambulatory patient storage, such as when the patient carries the pen injector for use during the day (e.g., on public transportation or while running). |
Last dose DA | Test DA for the last dose | 0.975 | Only the last Unit | 60 | 2.670 | The final dose is considered to be as important as any other dose and therefore it is appropriate to meet the same accuracy requirements. |
Thermal humid | Test DA after 40℃, 93%RH for 96 hours | 0.975 | 1 | 20 | 2.670 | This test is intended to measure the performance of a pen exposed to the same temperature as the dry heat test temperature but at a much higher humidity. |
Dose Accuracy Test Item | Test Environment | Samples for Resistance | Samples for Needle Fitting | Samples for Leakage Check |
Standard Atmosphere | 23°C±5°C | 10 | 20 | 10 |
Cold Atmosphere | 5°C±3°C | 10 | 20 | 10 |
Warm Atmosphere | 40°C±2°C | 10 | 20 | 10 |
Dry Heat | 70℃±2℃ | 10 | 20 | 10 |
Freezing | -40°C±3℃ | 10 | 20 | 10 |
Vibration | After vibration | 10 | 20 | 10 |
Thermal Humid | 40℃, 93%RH for 96 hours | 10 | 20 | 10 |